Series 4 DIGITAL FORCE GAUGES User’s Guide
Series 4 Digital Force Gauges User’s Guide 9 When set points are enabled, the following indicators are shown to the left of the primary rea
Series 4 Digital Force Gauges User’s Guide 10 Press ENTER to delete the value. To exit Delete mode, press DELETE again. Any number of readi
Series 4 Digital Force Gauges User’s Guide 11 or Click “Install”. 3. The next screen appears as follows: Click “Continue Any
Series 4 Digital Force Gauges User’s Guide 12 Select “No, not this time”, then click “Next”. 6. The next screen appears as follows: S
Series 4 Digital Force Gauges User’s Guide 13 8. The next, and final, screen appears as follows: Click “Finish”. The Mark-10 USB device i
Series 4 Digital Force Gauges User’s Guide 14 Any detected errors are reported back by means of the following error codes: *10 Illegal c
Series 4 Digital Force Gauges User’s Guide 15 9.6 I/O Connector Pin Diagram (female) Pin No. Description Input / Output 1 Signal Ground -
Series 4 Digital Force Gauges User’s Guide 16 2. Press DIRECTION to invert the display, if desired. ENTER to continue. The display will app
Series 4 Digital Force Gauges User’s Guide 17 If failed: 6. The following screen appears after the offsets have been calcula
Series 4 Digital Force Gauges User’s Guide 18 10. The display will appear as follows: Use the UP and DOWN keys to adjust the loa
Series 4 Digital Force Gauges User’s Guide 1 Thank you… Thank you for purchasing a Mark-10 Series 4 digital force gauge, designed for tens
Series 4 Digital Force Gauges User’s Guide 19 13. At the completion of the tension calibration, the display will appear as follows:
Series 4 Digital Force Gauges User’s Guide 20 11 OTHER SETTINGS 11.1 Automatic Shutoff The gauge may be configured to automatically power
Series 4 Digital Force Gauges User’s Guide 21 Press ENTER to modify the contrast. Select a value from 0 to 25, 25 producing the most contr
Series 4 Digital Force Gauges User’s Guide 22 12 SPECIFICATIONS 12.1 General Accuracy: ±0.2% of full scale ±1 digit Sampling rate: 3,
Series 4 Digital Force Gauges User’s Guide 23 12.2 Factory Settings Parameter Setting Set points Upper Disabled (defaults to 80% of fu
Series 4 Digital Force Gauges User’s Guide 24 12.3 Capacity, Resolution & Load Cell Deflection 12.4 Dimensions IN [MM] Thread Hex
Series 4 Digital Force Gauges User’s Guide 25 NOTES: .ELECTROMATIC Equipment Co., Inc., 600 Oakland Ave., Cedarhurst, NY 11516 – USATel: 800-645-
Series 4 Digital Force Gauges User’s Guide 2 1 OVERVIEW 1.1 List of included items Qty. Part No. Description M4-012 – M4-20 M4-50 – M4-
Series 4 Digital Force Gauges User’s Guide 3 accumulate in the sample during testing. Extra bodily protection should be worn if a destructiv
Series 4 Digital Force Gauges User’s Guide 4 3 MECHANICAL SETUP 3.1 Loading shaft orientation In order to accommodate a variety of testin
Series 4 Digital Force Gauges User’s Guide 5 operator and others in the vicinity. If using a grip or fixture from a supplier other than Mark
Series 4 Digital Force Gauges User’s Guide 6 kN – Kilonewton mN – Millinewton Note: not all gauge capacities measure in all
Series 4 Digital Force Gauges User’s Guide 7 4.3 Menu navigation basics Most of the gauge’s various functions and parameters are configured
Series 4 Digital Force Gauges User’s Guide 8 Two filters are available: Current Reading – Applies to the peak capture rate of the instrum