Quick Guide v. 3.0
Ultrasonic Thickness Gage
Advanced model
Seitenansicht 0
1 2 3 4 5 6 ... 12 13

Inhaltsverzeichnis

Seite 1 - Ultrasonic Thickness Gage

Quick Guide v. 3.0 Ultrasonic Thickness GageAdvanced model

Seite 2 - Quick Start

Statistics ModeStatistics ModeA statistical summary will appear on the display. Remove the lastmeasurement by pressing the (-) button. Press

Seite 3

10Outputting Stored ReadingsOutputting Stored ReadingsUSB mass storage - connect your PosiTector to a PC/Mac usingthe supplied USB cable to access and

Seite 4 - How to Measure

11Auto SYNCAllows the instrument to automatically synchronize withPosiTector.net when initially connected to an internet connectedPC runn

Seite 5 - Surface Conditions

12www.defelsko.com© DeFelsko Corporation USA 2012All Rights ReservedThis manual is copyrighted with all rights reserved and may not be reproduced or t

Seite 6

IntroductionIntroductionThe PosiTector UTG is a hand-held Ultrasonic Thickness Gagethat uses the non-destructive ultrasonic pulse-echo principle

Seite 7 - Cal Settings Menu

2To disconnect a probe from a body, power-down theinstrument and slide the plastic probe connectorhorizontally (in the direction of the arrow

Seite 8

3How to MeasureHow to MeasureThe PosiTector UTGM in multiple-echo mode determinesthickness by measuring the time between at least thre

Seite 9 - Setup Menu

4Surface ConditionsSurface ConditionsUltrasonic measurements are affected by the condition,roughness and contour of the surface to be tested. M

Seite 10 - (Advanced models only)

5Calibration, Verification and AdjustmentCalibration, Verification and AdjustmentThree steps ensure best accuracy…1.Calibration - typically done by th

Seite 11 - Memory Management

Cal Settings MenuCal Settings MenuThe PosiTector UTG is factory calibrated. But in order for it to takeaccurate thickness measurements of a particular

Seite 12 - Connect Menu

7Cal LockThicknessThe most common method of adjustment is to measure a sampleof known thickness. Select a reference standard of material asc

Seite 13

ScanNormally, the PosiTector UTG takes a single spot measurementat a rate of 6 readings per second while in contact with a surface.When the probe is